Ground Bond tester

TOS6210

TOS6210

380,000 (税込:418,000

  • Test current value: 6 A to 60 A AC / Resistance value: 0.001 Ω to 0.600 Ω
  • Voltage drop-based judgment function
  • Offset cancelling function
  • Stores 100 test conditions in memory
  • Incorporates test conditions into program
  • Contact check function
  • Equipped with standard GPIB and RS232C interfaces
  • Equipped with standard test lead (TL12-TOS)

LAN LXIUKCA
Login to myKIKUSUI / What is myKIKUSUI?
カート機能が動くまでの

Overview


Ground Bond tester supporting standard compliance tests up to 60A

While inheriting the basic performance and functions of its predecessor (TOS6200), such as a constant current driving system that provides current waveforms with little skew and high measurement accuracy, the TOS6210 tester extends the maximum test current from 30 A to 60 A, which is demanded by the new standard. In addition, the tester also lets you judge the acceptability of the device under test based on the drop in voltage, as required in the standard.

What’s more, you can preset test conditions of up to 20 different types of safety standards, such as those for information technology equip- ment, home appliances, medical devices, and measuring in- struments, in the memory on the main unit’s panel. A simple memory call operation allows you to set up a pro- tective earth or protective bonding continuity test as stipu- lated in UL and other relevant specifications including IEC and JIS standards. The tester also features a set of functions that meet the specific needs of testing personnel, such as an offset cancellation function and a memo function that allows you to input calibration dates, production numbers, and other test-related information and read the input information later via the GPIB or RS232C interface.

Specifications

Specifications

ZOOM

Output

Current setting range *1 *4 Resolution 6.0 to 62.0 A AC (With respect to resistance resulting in output power of the maximum rated Output or less and an output terminal voltage of 5.4 V or less)
Resolution0.1 A
Accuracy± (1% of setting + 0.4A)
Maximum rated output220 VA (at the output terminals)
Distortion2% or less (with respect to 0.1 Ω pure resistance load of 20 A or greater)
Frequency 50/60 Hz, sine wave (selectable)
Accuracy±200 ppm
Open terminal voltage6 Vrms or less
Output methodPWM switching

Output ammeter

Measurement rangeAC 0.0A to 66.0A
Resolution / Accuracy0.1A / ± (1% of reading + 0.4A)
ResponseMean value response/rms value display (response time: 200 ms)
Holding functionThe current measured at the end of test is held during the PASS or FAIL inteval

Output voltmeter

Measurement range0.00 to 6.00 V AC
Resolution / Accuracy0.01V / ± (1% of reading + 0.02V)
Offset cancel function0.00 to 5.40 V (Offset ON/OFF function provided)
ResponseMean value response/rms value display (response time: 200 ms)
Holding functionThe voltage measured at the end of test is held during the PASS or FAIL inteval

Ohmmeter *2

Measurement range0.001 to 0.600 Ω
Resolution / Accuracy0.001 Ω / ± (2% of reading + 0.003 Ω)
Offset cancel function0.000 to 0.600 Ω (Offset ON/OFF function provided)
Holding functionThe resistance measured at the end of test is held during the PASS or FAIL interval

Pass/fail judgement function *3

Resistance value-based judgement 

Window comparator system

  • If a resistance value equal to or greater than the upper reference value is detected, a FAIL determination is returned.
  • If a resistance value equal to or less than the lower reference value is detected, a FAIL determination is returned.
  • If a resistance value has been judged as FAIL, the tester shuts off the output and generates a FAIL signal.
  • If the set time elapses without abnormalities, the tester shuts off the output and generates a PASS signal.
Setting range for the upper reference value (UPPER)0.001 to 0.600 Ω
Setting range for the lower reference value (LOWER)0.001 to 0.600 Ω
Resolution0.001Ω
Judgment accuracy± (2% of UPPER + 0.003 Ω)
Sampled voltage value-based judgment 

Window comparator system

  • If a voltage value equal to or greater than the upper reference value is detected, a FAIL determination is returned.
  • If a voltage value equal to or less than the lower reference value is detected, a FAIL determination is returned.
  • If a resistance value has been judged as FAIL, the tester shuts off the output and generates a FAIL signal.
  • If the set time elapses without abnormalities, the tester shuts off the output and generates a PASS signal.
Setting range for the upper reference value (UPPER) *40.01 to 5.40 V
Setting range for the lower reference value (LOWER)0.01 to 5.40 V
Resolution0.01V
Judgment accuracy± (2% of setting + 0.05 V)
CalibrationCalibration is performed with the rms value of the sine wave, using a pure resistance load.
LEDPASSLights for approximately 0.2 sec when the measured value has been judged as PASS. It is lit continuously when the PASS holding time is set to HOLD.
UPPER FAILLights if a resistance or voltage value equal to or greater than the upper reference value is detected and judged FAIL.
LOWER FAILLights if a resistance or voltage value equal to or greater than the upper reference value is detected and judged FAIL.
Buzzer

The buzzer sounds for the pass holding time has been set if the measured value has been judged as PASS.

  • The buzzer sounds continuously under the following condition:
    The measured value has been judged as PASS when the PASS holding time is set to HOLD.
    The measured value has been judged as UPPER FAIL. The measured value has been judged as LOWER FAIL.
  • The buzzer volume for FAIL or PASS judgment are adjustable.
    Note that it cannot be adjusted individually since setting is shared with the setting for PASS.

Time

Test timeSelectable range0.3 to 999 s Timer ON/OFF function is available.
Accuracy± (100ppm of setting + 20ms)

Environment

Operating environmentIndoor use, overvoltage category II
Warranty rangeTemperature5° to 35°C
Humidity20 %rh to 80 %rh (non condensing)
Operating rangeTemperature0° to 40°C
Humidity20 %rh to 80 %rh (non condensing)
Storage rangeTemperature-20° to 70°C
Humidity90 % rh or less (no condensation)
AltitudeUp to 2000 m

Power requirement

Input voltage range85 to 250 V AC
Power consumptionAt no load (READY)60 VA or less
Rated load420 VA max.
Input frequency range47 Hz to 63 Hz
Insulation resistance *430M min. (500 Vdc), between AC line and chassis
Withstanding voltage1390 Vac (2 seconds), between AC line and chassis
Earth continuity25 Aac/ 0.1 Ω max

Safety *5 Conforms to the requirements of the following directive and standard.

SafetyLow Voltage Directive 2014/35/EU*5 EN 61010-1 (Class I, Pollution degree 2)

Electromagnetic compatibility (EMC) *5 *6 Conforms to the requirements of the following directive and standard.

Electromagnetic compatibility (EMC)EMC Directive 2014/30/EU EN 61326-1 (Class A) EN 55011 (Class A, Group 1) EN 61000-3-2 EN 61000-3-3 Under following conditions
  1. Used test leadwire (TL12-TOS) which is supplied.
  2. Used the shielded cable which length is less than three meters when the SIGNAL I/O is used.

Accessories

AccessoriesAC power cord: 1 piece, Test leadwire TL12-TOS: 1 set, Short bar: 2 pieces (These are inserted between the OUTPUT and SAMPLING terminals.), AC power fuse: 2 pieces (2, including one spare in the fuse holder), Operation manual: 1 copy

*1 Time limitation with respect to output. The heat radiation capacity at the output block of the tester is designed to be one-third of the rated output, accounting for size, weight, cost, and other factors. Always use the tester within the limitation values given below. Use of the tester beyond these limits will cause the temperature of the output block to rise excessively, potentially tripping the internal protection circuit. In this case, suspend testing for approximately 30 minutes, then press the STOP switch. When temperatures fall to normal levels, the tester will revert to ready status.

Ambient temperature t (°C)Test current I (A)Pause timeMaximum allowable continuous test time
t ≤ 40°40 < I ≤ 60Equal to or greater than the test time≤ 10 minutes
20 < I ≤ 40Equal to or greater than the test time≤ 30 minutes
I ≤ 20Not requiredContinuous output possible

*2 About ohmmeter’s response time. A resistance value is instantaneously obtained, calculated using the measured voltage and current values. The response time of the ohmmeter complies with the response times of the voltmeter and ammeter. *3 Resistance value-based and sampled voltage value-based judgments cannot be simultaneously conducted. *4 Limited by the maximum rated output and the output terminal voltage. The tester can be used within the range shown below. *5 Not applicable to custom order models. *6 Only on models that have CE marking on the panel.

Allowable range in which to determine the test current value and upper reference value


Pin Configuration for the SIGNAL I/O Connector

Pin Configuration for the SIGNAL I/O Connector

Product Catalog

Icon

    Operation Manuals

    TOS6200A/TOS6210
    Icon

      Please feel free to contact us with any questions or concerns.

      Download a Product Catalog View List
      Request a Quotation Contact Us
      close

      Search

      Entering the model number, capacity, etc.