Voltmeter
| Item | PLZ1005WH2 | PLZ2005WH2 | PLZ4005WH2 | PLZ12005WH2 | PLZ20005WH2 | |
|---|---|---|---|---|---|---|
| Display | 0.00 V to 1000.00 V | |||||
| Resolution | 10 mV | |||||
| Accuracy | ±(0.05 % of reading + 0.05 % of rating) | |||||
| Parallel operation | ±(0.1 % of reading + 0.1 % of rating) (TYP) | |||||
Ammeter
| Item | PLZ1005WH2 | PLZ2005WH2 | PLZ4005WH2 | PLZ12005WH2 | PLZ20005WH2 | |
|---|---|---|---|---|---|---|
| Display | 0.000 A to 20.000 A | 0.000 A to 40.000 A | 0.000 A to 80.000 A | 0.00 A to 240.00 A | 0.00 A to 400.00 A | |
| Resolution | 0.001 A | 0.001 A | 0.001 A | 0.01 A | 0.01 A | |
| Accuracy | ±(0.2 % of reading + 0.1 % of rating) | |||||
| Parallel operation | ±(0.4 % of reading + 0.2 % of rating) (TYP) | |||||
Power display
| Item | PLZ1005WH2 | PLZ2005WH2 | PLZ4005WH2 | PLZ12005WH2 | PLZ20005WH2 |
|---|---|---|---|---|---|
| Display | Displays the product of the voltmeter reading and ammeter reading | ||||
Measurement trigger
| Item | PLZ1005WH2 | PLZ2005WH2 | PLZ4005WH2 | PLZ12005WH2 | PLZ20005WH2 |
|---|---|---|---|---|---|
| Trigger Source | Immediate, BUS, DIGITAL2, MSync, TALink, LoadOff | ||||
| Trigger Count | 1 to 65536 | ||||
| Trigger Delay | 0.00000 s to 100.00000 s | ||||
| Interval | Disable/Enable | ||||
| Interval Time | 0.00001 s to 3600 s | ||||
| Sense Aperture | 0.00001 s to 1.00000 s | ||||