Dimensions / Weight
W | H | D | |
---|---|---|---|
Dimensions | 224 mm (8.82 inch) | 113 mm (4.45 inch) | 373 mm (14.69 inch) |
Digit Multimeter
DME1600
DME1600
134,000円 (税込:147,400円)
- Resolution : 6 1/2 digit
- Display : 5 x 7 dot matrix VFD, dual display with 3-color
- Basic measurement function
- DC voltage : 0.1 V, 1 V, 10 V, 100 V, 1000 V
- AC voltage : 0.1 V, 1 V, 10 V, 100 V, 750 V
- DC current : 10 mA, 100 mA, 1 A, 3 A
- AC current : 1 A, 3 A
- 2-wire / 4-wire resistance : 100 Ω, 1 kΩ, 10 kΩ, 100 kΩ, 1 MΩ, 10 MΩ, 100 MΩ
- Frequency : 3 Hz to 300 kHz
- Continuity test
- Diode test
- Temperature test
- Built-in USB Interface (GPIB Interface※: selected model)
*Model with GPIB Interface : DME1600GC
Overview
Essential device of “Electronic Measurement”
Supporting basic measurement with variety of options
The DME1600 is a digital multi-meter with a resolution of 6 1/2 digit. It can be performed up to 2000 times per second at the setting condition of 4 1/2 digit as fastest measurement, and it can measures 50 times per second when it is set for the 6 1/2 digit. The DME1600 offers fully function of measurement for the voltage, current, resistance, frequency and temperature which can be used various application of measurement and evaluation in design, development and debugging of electronics devices. The DME1600 provides USB and GPIB interface as standard feature for automated measurement besides manual operation. Furthermore, the DME1600 offers wide range of options such as 20-channel multi-point scanner card supporting the basic measurement.
Specifications
DC Characteristics
Accuracy
- ±(% of reading + % of range)
- The specifications are for the following conditions: 6 1/2 digit resolution, minimum two-hour warm up, and auto trigger mode.
- Use the null function for the 2-wire / 4-wire resistance measurement method.
DC voltage
Range | Resolution | Input resistance | 1 year (23 °C ± 5 °C) |
---|---|---|---|
100.0000 mV | 0.1 μV | > 10 GΩ | 0.0050+0.0035 |
1.000000 V | 1.0 μV | > 10 GΩ | 0.0040+0.0007 |
10.00000 V | 10 μV | > 10 GΩ | 0.0035+0.0005 |
100.0000 V | 100 μV | 10 MΩ | 0.0045+0.0006 |
1000.000 V | 1 mV | 10 MΩ | 0.0045+0.0010 |
DC current
Range | Resolution | Input resistance | 1 year (23 °C ± 5 °C) |
---|---|---|---|
10.00000 mA | 10 nA | 5.1 Ω | 0.050+0.020 |
100.0000 mA | 100 nA | 5.1 Ω | 0.050+0.005 |
1.000000 A | 1 μA | 0.1 Ω | 0.100+0.010 |
3.000000 A | 10 μA | 0.1 Ω | 0.120+0.020 |
Resistance
Range | Resolution | Input resistance | 1 year (23 °C ± 5 °C) |
---|---|---|---|
100.0000 Ω | 100 μΩ | 1 mA | 0.010+0.004 |
1.000000 kΩ | 1 mΩ | 1 mA | 0.010+0.001 |
10.00000 kΩ | 10 mΩ | 100 μA | 0.010+0.001 |
100.0000 kΩ | 100 mΩ | 10 μA | 0.010+0.001 |
1.000000 MΩ | 1 Ω | 5 μA | 0.010+0.001 |
10.00000 MΩ | 10 Ω | 500 nA | 0.040+0.001 |
100.0000 MΩ | 100 Ω | 500 nA||10 MΩ | 0.800+0.010 |
Diode test
Range | Resolution | Test current | 1 year (23 °C ± 5 °C) |
---|---|---|---|
1.0000 V | 10 μV | 1 mA | 0.010+0.020 |
Continuity test
Range | Resolution | Test current | 1 year (23 °C ± 5 °C) |
---|---|---|---|
1 kΩ | 10 mΩ | 1 mA | 0.010+0.030 |
Measurement characteristics
Item | Specifications |
---|---|
DC voltage measurement : Overrange | Permits voltages that are up to 20 % over the range except when the 1000 V range is in use |
DC voltage measurement : Input bias current | Less than 30 pA (at 25 °C) |
DC voltage measurement : Input voltage protection | 1000 V for all ranges |
DC current measurement : Overrange | Permits currents that are up to 20 % over the range except when the 3 A range is in use |
Resistance measurement : Maximum test lead resistance | 10 Ω (100 Ω range) |
100 Ω (1 kΩ range) | |
1 kΩ (other ranges) | |
Resistance measurement : Input voltage protection | 1000 V for all ranges |
AC Characteristics
Accuracy
- ±(% of reading + % of range)
- The specifications are for the following conditions: 6 1/2 digit resolution, minimum two-hour warm up, and slow AC filter (3 Hz to 300 kHz bandwidth).
- Measured using a sine wave input whose amplitude is greater than 5% of range.
- For the input range from 1 % to 5 %, add 0.1 % of the range (when it is less than 50kHz) or adding 0.13 % of the range (when it is from 50 kHz to 100 kHz)
AC voltage (true rms value)
Range | Resolution | Frequency | 1 year (23 °C ± 5 °C) |
---|---|---|---|
100.0000 mV | 0.1 μV | 3 Hz to 5 Hz | 1.00+0.04 |
5 Hz to 10 Hz | 0.35+0.04 | ||
10 Hz to 20 kHz | 0.06+0.04 | ||
20 kHz to 50 kHz | 0.12+0.05 | ||
50 kHz to 100 kHz | 0.60+0.08 | ||
100 kHz to 300 kHz | 4.00+0.50 | ||
1.000000 V to 750.000 V | 1.0 μV to 1 mV | 3 Hz to 5 Hz | 1.00+0.03 |
5 Hz to 10 Hz | 0.35+0.03 | ||
10 Hz to 20 kHz | 0.06+0.03 | ||
20 kHz to 50 kHz | 0.12+0.05 | ||
50 kHz to 100 kHz | 0.60+0.08 | ||
100 kHz to 300 kHz | 4.00+0.50 |
AC current (true rms value)
Range | Resolution | Frequency | 1 year (23 °C ± 5 °C) |
---|---|---|---|
1.000000 A | 1 μA | 3 Hz to 5 Hz | 1.00+0.04 |
5 Hz to 10 Hz | 0.30+0.04 | ||
10 Hz to 5 kHz | 0.10+0.04 | ||
3.000000 A | 10 μA | 3 Hz to 5 Hz | 1.10+0.06 |
5 Hz to 10 Hz | 0.35+0.06 | ||
10 Hz to 5 kHz | 0.15+0.06 |
Measurement characteristics
Item | Specifications |
---|---|
Overrange | Capable 20 % of the over rage excluding “750 V RMS range” |
Measurement frequency | 750 V RMS Range is limited to 100 kHz |
Frequency and period characteristics
Accuracy
- ±(% of reading)
- The specifications are for the following conditions: 6 1/2 digit resolution and minimumtwo-hour warm up.
Range | Frequency | 1 year (23 °C ± 5 °C) |
---|---|---|
100 mVrms to 750 Vrms | 3 Hz to 5 Hz | 0.10 |
5 Hz to 10 Hz | 0.05 | |
10 Hz to 40 Hz | 0.03 | |
40 Hz to 300 kHz | 0.01 |
Measurement characteristics
Item | Specifications |
---|---|
Overrange | Permits voltages that are up to 20 % over the range except when the 750 Vrms range is in use |
Measurement frequency | 750 V RMS Range is limited to 100 kHz |
General specifications
Item | Specifications |
---|---|
Input voltage range | 100 Vac/120 Vac/220 Vac/240 Vac ± 10 % (single phase) |
Input frequency range | 50 Hz/60 Hz±10 % |
Power consumption | 25 VAmax |
Operating temperature range | 0 °C to 50 °C |
Operating humidity range | Up to 80 % rh (0 °C to 31 °C, non condensing) |
Storage temperature range | -40 °C to 70 °C |
Operating altitude | Up to 2000 m |
Dimensions/Weight | 224(8.82)W×113(4.45)H×373(14.69)D mm(inch) Approx. 3.7 kg (8.2 lb) |
Interface | USB 2.0, GPIB (factory option) |
Safety *1 | Complies with the requirements of the following directive and standard. Low Voltage Directive 2014/35/EU EN 61010-1 (Class I, Pollution degree 2), EN 61010-2-030 |
Electromagnetic compatibility (EMC) *1 *2 | Complies with the requirements of the following directive and standard. EMC Directive 2014/30/EU EN 61326-1 (Class B ) EN 55011 (Class B, Group 1) EN 61000-3-2, EN 61000-3-3 |
Accessories | “Power cord” 1 pc. (with three-pronged plug), “Standard test leads” (1 red, 1 black), “USB cable” 1pc., “Fuse”(spare) 1pc., “CD-ROM”*1pc., “Packing list,safety precautions” (1 English, 1 Japanese) *Contains the user’s manual and the remote interface manual. |
*1 Does not apply to specially made or modified DME1600s. *2 Limited to products that have the CE/UKCA mark on their panels. *3 This is a Class I equipment. Be sure to ground this product’s protective conductor terminal. The safety of this product is only guaranteed when the product is properly grounded. *4 Pollution is addition of foreign matter (solid, liquid or gaseous) that may produce a reduction of dielectric strength or surface resistivity. Pollution Degree 2 assumes that only non-conductive pollution will occur except for an occasional temporary conductivity caused by condensation. *5 This is a Class A equipment. This product is intended for use in an industrial environment. This product may cause interference if used in residential areas. Such use must be avoided unless the user takes special measures to reduce electromagnetic emissions to prevent interference to the reception of radio and television broadcasts. *6 This is a Group 1 equipment. This product does not generate and/or use intentionally radio-frequency energy, in the form of electromagnetic radiation, inductive and/or capacitive coupling, for the treatment of material or inspection/analysis purpose. *7 Contains the User’s Manual and the Remote Interface Manual.