SD005-KHA (Harmonics Explorer) – Update

Update Module for Harmonics Explorer

  • This is the latest update module for the “Harmonics Explorer”.
  • To use this module, you will also need to upgrade the KHA1000 main unit.
  • In order to install this module, the full version of SD005-KHA Harmonics Explore must already be installed.
  • Please be aware that this software is not compatible with Windows XP and previous operating systems.

Ver.

3.9.1

Last update

2023/8/8

Configuration

  • Windows 10 (x86/x64)
  • Windows 11

Download

Release Note

VER 3.9.1
2023-7-04

HarmoCapture / HA File Analyzer

Resolved issues
The description of the standard for measurement techniques that conforms to JIS C61000-3-2(2019) has been revised from JIS C61000-4-7(2007) to IEC61000-4-7 Ed2.1(2009).
VER 3.9.0
2021-11-26

HarmoCapture / HA File Analyzer / Vf File Analyzer

New Feature
IEC61000-3-2 Ed4.0 of HA limitation standard was changed to IEC61000-3-2 Ed5.0.
New Feature
JIS C 61000-3-2 2011 of HA limitation standard was changed to JIS C 61000-3-2 2019.
New Feature
IEC61000-3-3 Ed3.0 of Vf flicker limitation standard was changed to IEC61000-3-3 Ed3.1.
VER 3.8.0
2021-4-08

HarmoCapture / Vf File Analyzer

New Feature
This version operates also in AC power supply PCR-WEA series.
Resolved issues
There was a problem whose output ON/OFF of AC power supply in the instrument menu was not effective.
Updated
The front name of radio buttons in voltage range of AC power supply has been replaced Low-High from 100V-200V.
Resolved issues
While not opening the result file by Vf File Analyzer, there was a problem to which an exception error occurs, when Segment information was clicked.
VER3.7.1
2018-5-10

Vf File Analyzer

Resolved issues
When IEC61000-4-15 is Ed2.0, IEC61000-3-3 is Ed3.0. However, 3-3 Ed2.0 was a mistake.
VER 3.7.0
2017-5-26

HarmoCapture / HA File Analyzer / Vf File Analyzer

Updated
For HA limitation standards, "IEC61000-3-2 Ed4.0" have been added.
Updated
For Flicker limitation standards, "IEC61000-3-3 Ed3.0" have been added.
VER 3.6.5
2016-9-02

HarmoCapture

Resolved issues
There was a problem which the voltage range of PCR-LE could not change. This problem occured only in Ver3.6.4
VER 3.6.4
2016-4-05

HarmoCapture

Resolved issues
There was a problem which does not conform to IEC61000-4-15 Ed2.0. The steady state and d measurement algorithm were changed.
VER 3.6.3
2015-2-03

HA File Analyzer

Resolved issues
There was a problem which had not been taking the magnification of a current range into consideration by text conversion of a V/I waveform.
VER 3.6.2
2014-9-02

HarmoCapture

Resolved issues
There was a problem that the suitable value had not been shown in dialog of "Check the line impedance".
Resolved issues
When Key input to a NumericUpDown control, there was a problem that nominal voltage and frequency had returned to a former setup at click of Send-button.
Updated
Change to read setting value from KHA at the time of the first HA-Vf change after boot of HarmoCapture.
VER 3.6.1
2014-7-01

Vf File Analyzer

Resolved issues
Vf File Analyzer Ver3.6.0 had a problem which could not read a result file created by KHA1000 Ver1.73 or earlier.
VER 3.6.0
2013-12-06

HarmoCapture

New Feature
For HA limitation standards "JIS C 61000-3-2 2011" have been added.
Improvement
Measurement time was extended to 9600 seconds.
New Feature
Corresponded to remote control of LIN-JF series.
VER 3.5.5
2012-9-14

HarmoCapture

New Feature
AC power supply PCR-LE series have been added to the operation.
VER 3.5.4
2011-10-18

HarmoCapture

Resolved issues
There was a problem that send progress bar was not zero clear even if changed radio button of flicker measurement technic "IEC61000-4-15 Ed1.1" and "IEC61000-4-15 Ed2.0".
VER 3.5.3
2011-5-16

HarmoCapture / HA File Analyzer / Vf File Analyzer

Improvement
English contents and user's manual was updated.
Improvement
English contents and user's manual was updated.
Improvement
English contents and user's manual was updated.
VER 3.5.2
2011-2-21

HarmoCapture

Resolved issues
When operated HA -> Vf -> HA, there was a problem which the standard group name had changed to "Standard 61000-3-3 / 4-15".
VER 3.5.1
2011-1-25

HarmoCapture

Resolved issues
When setting files except JIS2005 were opened, there was a problem which an error occurs.
Resolved issues
For Flicker measurement technic "IEC61000-4-15 Ed1.1" and "IEC61000-4-15 Ed2.0" can be saved now in vs file.
Resolved issues
When a hs file before Ver3.4.5 was read, there was a problem which a standard changes.
VER 3.5.0
2010-12-16

HarmoCapture / HA File Analyzer / Vf File Analyzer

Improvement
For HA limitation standards, "IEC61000-3-2 Ed3.0 A2" and measurement technic "IEC61000-4-7 Ed2.0 A1" have been added.
New Feature
Reference impedance is now selectable to "Use" or "Un-use" when JIS2005 selected.
Improvement
For Flicker limitation standards, "IEC61000-3-3 Ed2.0" and measurement technic "IEC61000-4-15 Ed2.0" have been added.
New Feature
Confirmation of AC power source have been added. This is applied to screen display and reports.
Improvement
Only when HA limitation standard selected IEC61000-3-2 Ed3.0 A2, calculation denominator of a repeatability is the limit value. Less than 5% is Pass.
New Feature
Alias Standard of measurement technic for Ed1.1 and Ed2.0 are added.
VER 3.4.5
2010-9-15

HarmoCapture / HA File Analyzer / Vf File Analyzer

New Feature
IEC61000-3-2 Ed3.0 A2 / 61000-4-7 Ed2.0 A1 and IEC61000-3-2 Ed3.0 A2 / 61000-4-7 Ed1.0 have been added to HA. By this addition, the latest standards and other standards are now put into the radio-buttons and a combox respectively.
New Feature
Only when Class C is selected under IEC Ed3.0A2 / Ed2.0A1 or IEC Ed3.0A2 / Ed1.0, 3rd/5th/current-waveform are now displayed at applied limitation values.
Improvement
The standard selection IEC Ed1.2/Ed1.1 is changed to IEC Ed2.0/Ed1.1 for Vf.
New Feature
A description figure of current waveform requirement is now drawn at V/I waveform when a result file where the applied limit values are 3rd/5th/current-waveform in Class-C. This is applied to screen display and reports.
New Feature
Because "Use Alias Standard" and related items are added, an arbitrary standard name is now able to print in reports.
New Feature
The 2D harmonic current auto-scale is now selectable to "Meas Value" or "Meas and Limit". In "Meas and Limit", the appropriate scale is calculated from measurement values and limit values, therefore the printing problem that the limit values were too high to print has been fixed.
New Feature
Because "Use Alias Standard" and related items are added, an arbitrary standard name is now able to print in reports.
VER 3.4.4
2009-12-03

HarmoCapture

Resolved issues
There was a problem that the final judgement on report could be FAIL in spite of test result passed.
VER 3.4.3
2009-10-15

HarmoCapture / HA File Analyzer

Improvement
The setting screen is now greyed out while testing so that parameter settings cannot be altered.
Resolved issues
There was a problem that the max apparent power value in the report was shown as the max value during test period.
VER 3.4.2
2009-7-10

HarmoCapture / HA File Analyzer / Vf File Analyzer

Improvement
There was a problem that test reports under IEC61000-3-2 Ed3.0 incorrectly indicated Limit Over(s) values even if other than Class A.
Improvement
At IEC61000-3-3 manual switching d-measurement, there was a problem that the remaining time display was extremely wrong when starting the test without sending test condition after the previous measurement was once finished.
Improvement
There was a problem that the file size of report pdf could be extemely big. It has no side effect on printout itself.
Improvement
There was a problem that the V/I waveform printouts indicated fixed 0A values when the current was minimal such as when +Peak is 0A and -Peak is 0.05A.
Improvement
There was a problem that test reports under IEC61000-3-2 Ed3.0 incorrectly indicated Limit Over(s) values even if other than Class A.
Improvement
When testing at other than 230V/50Hz, the report incorrectly printed out EN standard name in spite of the specification not printing it.
Improvement
There was a problem that the "*3" and "*4" remarks on the IEC61000-3-2 Ed3.0 report could be incorrecly shown by the side effect of previous test.
Improvement
At the Vf Report Voltage Fluctuation graphs, there was a problem that the arrows indicating the start-point of dc, dmax, and dt, could not be printed out.
VER 3.4.1

HarmoCapture / HA File Analyzer / VF File Analyzer

Resolved issues
Corrected a typo found in generated reports (Fluction → Fluctuation).
Resolved issues
There was a problem that report pdf could be unnecesarily huge.
VER 3.4.0
2008-11-21

HA File Analyzer / VF File Analyzer

Improvement
Footer option has been added to the report functionality.
VER 3.3.1
2008-6-20

HarmoCapture / HA File Analyzer / Vf File Analyzer

Improvement
For HA test standards, "IEC61000-3-2 Ed3.0/4-7 Ed2.0" and "IEC61000-3-2 Ed3.0/4-7 Ed1.0" have been added.
Improvement
The "Limit Over" item has been added to test reports of "IEC61000-3-2 Ed3.0".
Improvement
The Vf test standard "IEC61000-3-3 Ed1.1" has been changed to "Ed1.2".
Improvement
When send operation finished, a message "Send was completed" is now displayed beneath the send progress bar.
Improvement
A message "remaining time" is now displayed beneath the test progress bar.
Improvement
For the test progress bar at Vf Manual Switching operation, it now represents 100% for each segment. Furthermore, "Segment (measuring)/(total)" is now displayed.
Improvement
At Vf Manual Switching, a dialogue box prompt is now shown when you wait for the next measurement, in order to increase visiblity.
Improvement
It is now compatibe with the added standard "IEC61000-3-2 Ed3.0"
Improvement
The flicker list view has been added. Furthermore the flicker list page is also added to reports.
VER 3.2.1.2
2008-4-11

VF File Analyzer

Resolved issues
In Vf reports of Pst Auto Switching mode, measurement time and count of Manual Switching were incorrectly used.
VER 3.2.1
2008-3-27

HarmoCapture / VF File Analyzer

Improvement
Limitation judgement (Judge Factor) are now selectable from pst, plt, dc, dmax, and d(t)>3.3%.
VER 3.1.0
2007-7-23

HarmoCapture / HA File Analyzer / VF File Analyzer

Improvement
As the item name "Comment" on the menu and toolbars is not comprehensive, it has been changed to "Report Setting".
Resolved issues
When openning a result file captured under Class-C by using KHA1000 mains VER1.20 or earlier versions, there were errors on current values. (This fault is found on V/I current waveform graph only.)
Improvement
As the item name "Comment" on the menu and toolbars is not comprehensive, it has been changed to "Report Setting".
Improvement
2D Harmonics Graph is now printable on the report.
Improvement
V/I Waveform Graph is now printable on the report.
Improvement
Harmonic Trend Graph is now printable on the report.
Improvement
Current Trend Graph is now printable on the report.
Improvement
THC Trend Graph is now printable on the report.
Improvement
Setting List is now printable on the report.
Improvement
Repeatability confirmation list is now printable on the report.
Improvement
As the item name "Comment" on the menu and toolbars is not comprehensive, it has been changed to "Report Setting".
Improvement
Voltage Fluctuation Graph of the selected segment is now printable on the report.
Improvement
Zoom-In graph on 'dc' is now printable on the report.
Improvement
Zoom-In graph on 'dmax' is now printable on the report.
Improvement
Zoom-In graph on 'd(t)>3.3%' is now printable on the report.
Improvement
Segment Information list is now printable on the report.
Improvement
Setting List is now printable on the report.
VER 3.0.0
2007-4-13

HarmoCapture / HA File Analyzer / VF File Analyzer

Improvement
An overwrite prompt dialog has been added when generating PDF files.
Improvement
Now compatible with the high-precision updagade of the KHA1000 mains.
Resolved issues
There was a problem that num of frames and their time slice are mismatched on Trend Graph's test time.
Resolved issues
There was a pronlem that the red and yellow colouring of 2D harmonic bar-graphs was incorrect.
Resolved issues
There was a problem that the display of [Ignore]/[Not Ignore] options was incorrectly reversed.
Improvement
Decimal display formats for Over-100% and Over-Margin have been changed so that they now match with the KHA1000 mains.
Improvement
Separator format for text save is now selectable.
Improvement
An overwrite prompt dialog has been added when generating PDF or text files.
Improvement
Now compatible with the high-precision updagade of the KHA1000 mains.
New Feature
This application is newly added.
VER 2.0.0
2006-11-06

HarmoCapture / HA File Analyzer

Resolved issues
There was a probem that PCs equiping OHCI USB host and any Windows2000 environment could not acquire measurement data through the USB interface.
Resolved issues
There was a problem that PCR-L could not be controlled.
Improvement
Upon report printing, it now prints the result file of the test immediately before. The functionality to print an arbitrary file has been moved to [File]->[Select file and print report...] menu.
Improvement
Upon class selection, "61000-3-2/4-7 IEC Ed2.2/Ed1.0", and "61000-3-2/4-7 JIS 2005/Ed1.0" have been added.
New Feature
This application is newly added.
VER 1.0.0
2006-3-22

HarmoCapture

New Feature
The first release