Update Module for BPChecker3000 (V1/V2)
This file is the latest updater of “BPChecker3000” for PFX2512/PFX2532. If you currently use the Ver 1, and to use the OP03-PFX, you will need to apply this update. Also, if you want to install this module , the product version “BPChecker3000” must have already been installed, in addition, you will need to update the firmware of PFX2512/PFX2532. Please refer to BPC3000_V2_UpdateGuide_E.pdf for details .
Caution: Please read before updating.
(1) This update does not support custom-made products.If your Model ID is in the 8000 series, your product is a custom-made item; please contact us regarding this update.
(2) The following notes apply if you are updating Ver 1 to Ver 2.
- When you’re using Ver1.0.0.0-Ver1.4.9.0, this update is applied. For a custom-made product, Ver1.5.0.0.-Ver1.9.9.9 can’t use this update.
- As a result of functional additions in Ver 2, the test condition file was changed from the data format used in Ver 1. When a test condition file created with Ver 1 is opened with Ver 2, the data is automatically updated for Ver 2. As such, new test conditions created with Ver 2 and test conditions created with Ver 1 that have been opened and saved with Ver 2 cannot be used with Ver 1. We recommend that you back up test conditions created with Ver 1 in a separate folder or the like.
- Ver 2 is designed so that test conditions acquired with Ver 1 can be viewed with Graphviewer of Ver 2, but problems may occur depending on the content of test data, the PC performance, and so on. If a problem occurs such as not being able to open test conditions of Ver 1 after updating to Ver 2, it is possible to revert temporarily to Ver 1. See BPC3000_V2_UpdateGuide_E.pdf.
- Ver 2 does not support Windows XP.
- The software environmental settings (graph axis, color, etc.) are passed on from Ver 1, but some settings may be reset to default values.
Ver.
2.0.9.01
Last update
2026/7/29
Configuration
- Windows10 (x86/x64) *About Windows10 support
- Windows11
- LAN
Download
Release Notes
VER 2.0.9.01
2026-07-29
PFX firmware / BPC3000
- Improvement Added support for additional model IDs.
- New Feature Added LAN communication capability to the temperature chamber interface.
VER 2.0.8.02
2025-03-04
PFX firmware / BPC3000
- Resolved issues Fixed the issue where a test does not complete at the set minimum voltage when the test is performed using an I-V characteristics profile created with the Test Condition Editor.
- Resolved issues Fixed the issue where an "element is out of effective range" error occurs when a BPCL3 file is loaded with the Graph Viewer.
- Improvement This version operates in Windows 11.
VER 2.0.7.00
2023-02-14
PFX firmware
- Resolved issues Fixed a problem that may not work with CV charging when charging CC-CV using a battery with high impedance.
- Resolved issues Fixed a problem that only the maximum cell voltage is not displayed in the test result when OP02-PFX is installed.
- Resolved issues Fixed a problem that an incorrect mode is displayed (recorded) when operating in cell CV mode.
VER 2.0.6.02
2021-1-20
PFX firmware / BPC3000
- Resolved issues Fixed a problem that activated the overvoltage protection (OVP) upon completion of the charge in combination with PWR-01 series.
- Resolved issues Fixed a problem when the test was paused and resumed in the last profile, only one profile was excecuted to complete the test regardless of the resuming position.
- Resolved issues Fixed a problem that decreased the detection accuracy of the overcharge capacity protection (OAH) (the OAH detection value (capacity: Ah) was rounded down to the nearest integer).
- Improvement Fixed a problem that required the time to switch profiles under the test condition with many profiles depending on the PC environment.
VER 2.0.5.00
2020-9-3
PFX firmware
- Resolved issues Fixed a problem that caused an error due to false detection of the option board when the option board was not inserted.
VER 2.0.4.00
2018-6-1
BPC3000
- New Feature Added support for model Ids that combine power supplies (PWR-01) and electronic loads (PLZ-5W).
- New Feature Added the “Save the data for Graph Viewer” option to TestExecutive. When the check box is cleared, memory consumption is suppressed, which makes long-term profile operation possible.
- Improvement Improved the memory usage algorithm and processing speed when a profile is run over a long time period on TestExecutive.
- Resolved issues Fixed a problem that was not applying selections correctly when a profile with the Delta Time of Data Record Method set to Fix Time is edited on Test Condition Editor.
VER 2.0.3.00
2017-12-1
PFX firmware / BPC3000
- Resolved issues Fixed a problem that was displaying an incorrect error code for electronic load alarms.
- Resolved issues Fixed a problem that the profile template created before V 2.01 is not displayed.
- Resolved issues Fixed a problem that was resetting the unbalanced margin display on the profile screen.
VER 2.0.2.12
2017-1-23
PFX firmware / BPC3000
- Resolved issues Fixed a problem that was always causing the test to stop with an OTP when I-V characteristic Prof was executed immediately after starting the PFX. (This problem could be reproduced only when a thermometer was not connected to the PFX.)
- New Feature Added support for the ID7306 model.
- Improvement Improved the PFX search algorithm on TestExecutive and IoConfig.
- Resolved issues Refined the model ID file loading process.
- Resolved issues Fixed a problem that was always causing the cell CV to operate during CC-CV charge Prof or pattern Prof and preventing current from flowing when all cell voltage measurements were turned off when an OP board was installed.
- Resolved issues Fixed a problem that causing an exception when no channels could be connected during TestExecutive startup.
VER 2.0.1.03
2016-7-29
PFX firmware / BPC3000
- Resolved issues Fixed a problem that was causing the cell CV function to operate when a given condition was met even when there was no OP board.
- Resolved issues Fixed a problem that was preventing programs from proceeding to the next loop if “RestTemp (LastStep)” is enabled in a pattern profile or I-V characteristic profile when there was no thermometer connected (e.g., OPEN).
- Improvement Redesigned the new text output format. (The two-row column header was changed to one.)
- Improvement Redesigned the GraphViewer text export function so that the elapsed time column is changed to the total time column.
- New Feature Added a preset timeout function for temperature chamber synchronization.
- Improvement Changed the supported KI-VISA version to v5.5.0. (To use temperature chambers, update the KI-VISA version to v5.5.0 or later.)
VER 2.0.0.38
2016-1-6
Global / BPC3000
- New Feature Added a cell CV function.
- New Feature Voltage measurement of up to 72 cells is possible by connecting SL01-PFXs.
- Improvement Drastically improved the operability of Graph Viewer. [Description] • Added an auto creation wizard for I-V characteristic graphs. • Printing can be executed from the shortcut menu. • Improved the operability of moving and zooming data lines. • Redesigned the pane configuration. • Redesigned the shortcut menu. • Redesigned button arrangement.
- Improvement Drastically improved the print content of Graph Viewer. [Description] • Improved the smudging of graphs during printing. • Redesigned grayscale printing. • Redesigned the print format of test condition and the like.
Ver 1.4.9.00
VER 1.4.9.00
2015-5-27
PFX2512 / BPC3000
- Resolved issues Fixed a problem that was causing the start of the next test to be held until the stop time expires even when the test itself has stopped earlier.
- Resolved issues Expanded the measurable temperature of the PFX thermometer from down to -40°C to down to -50°C.
- Improvement Updated the model ID profile to suppress the disparity between electronic load devices.
VER 1.4.8.01
2014-10-6
BPC3000
- Resolved issues Fixed a problem with the IT functionality not operating correctly.
- Resolved issues Fixed a problem that was preventing the 0.1 mV and 0.1 mA accuracies from being achieved in high-speed sampling mode.
- Improvement Fixed so that VISA resources LAN and GPIB can be selected in addition to serial.
- Resolved issues Fixed other problems.
VER 1.4.7.97
2013-7-9
BPC3000
- Resolved issues Changed the description of the chamber connection method in the operation manual.
VER 1.4.7.94
2013-6-6
BPC3000
- Improvement Improved the update feature so that users using special-order products can also continue to use their environment
VER 1.4.6.83
2013-3-10
PFX2512 / Test Executive / Test Condition Editor / GraphViewer
- Resolved issues Fixed the overshoot that was occurring the first time the PAT output was turned on.
- Resolved issues Fixed an error in the default current control method of pattern charge/discharge.
- Improvement Added a default time-based monitoring feature for pattern charge/discharge.
- Resolved issues Fixed a problem in the handling of the limit current in CP charge/discharge.
- Improvement Added support for 2-port LAN.
- Improvement Added a feature that saves the manually input IP address.
- Resolved issues Fixed an error in the CV value for CC-CV charge and CC-CP discharge in the profile master.
- Improvement Optimized the profile master.
- Resolved issues Fixed a problem that was causing incorrect processing when the edit button was processed if the profile master was empty.
- Resolved issues !!!DELETE!!!
- Resolved issues Fixed a problem that was causing the extension of life text file names to be changed.
VER 1.4.3.57
2013-1-31
Test Executive
- Resolved issues Fixed a problem that was preventing current graphs greater than or equal to 50 A from being displayed.
VER 1.4.2.56
2013-1-8
BPC3000 / Test Condition Editor / IO Config / Test Executive
- Improvement Added model ID 7119.
- Resolved issues Fixed a problem that was causing the application to close when model ID 7114 was selected.
- Improvement Prohibited the voltage range to be switched from IoConfig.
- Resolved issues Fixed a problem in which the extension of life result file names was lacking the dot.
- Improvement Changed the default chamber humidity setting to -1.
- Resolved issues Fixed a problem that was causing the low voltage range to return to high when a re-test was performed.
- Resolved issues Fixed an error in the updating of the test condition grid of TestExecutive.
VER 1.4.1.36
2012-11-02
BPC3000
- Resolved issues Fixed a problem that occurred when the chamber humidity was not set.
- Resolved issues Added model ID 7114.
VER 1.4.0.35
2012-10-22
BPC3000 / PFX2512
- Improvement Added an English help document.
- Resolved issues Fixed a problem in the trigger specification of pattern profiles.
VER 1.3.6.33
2012-8-22
GraphViewer / Test Condition Editor
- Resolved issues Fixed a problem that failed to show two profiles in the maximum and minimum displays of the cell voltage.
- Resolved issues Fixed a problem that was causing the time to drift when the cell voltage was expanded time-wise.
- Resolved issues Fixed a problem with the VI cursor movement.
- Resolved issues Fixed a problem with the total number of steps in patterns.
VER 1.3.3.28
2012-7-20
GraphViewer / PFX2512
- Resolved issues Fixed a problem with the processing of the Next button when option board data was displayed.
- Resolved issues Fixed a problem that was causing an exception when a BPCG3 file was opened directly and the content was zoomed.
- Resolved issues Fixed a problem that was showing abnormal values in the result data when the cell was excluded from measurements.
- Resolved issues Fixed a problem that was causing pattern charge/discharge execution times to be reduced in 60 Hz regions.
VER 1.3.2.26
2012-6-21
Test Condition Editor / GraphViewer
- Resolved issues Fixed a problem with the power limit when test condition files created on a special-order power supply model was being re-edited.
- Resolved issues Fixed a problem with the processing of the Next button when option board data was displayed.
VER 1.3.0.20
2012-4-24
Test Executive / PFX2512
- Resolved issues Fixed a problem that was causing an alarm when tests were executed without configuring the option board.
- Improvement Ver1.0.2 Changed the humidity calibration point.
VER 1.1.0.12
2012-3-27
Test Condition Editor / GraphViewer
- Improvement Added a profile master check and repair feature.
- Improvement Changed so that the X-axis can be set for each graph.