I’m controlling TOS5200 via RS232C from a PLC. When using the FETC:VOLT? command to read the voltage, why do I only receive CR LF?

The FETC command will only return a value if the MEAS/READ command was performed during the test. Please perform the MEAS/READ command during the test and then use the FETC command to check the measurement results. To obtain test results, use the RES? command.