When measuring resistance using the ANALOG OUT terminals on TOS7200, how long should I wait after starting the test?

When a capacitive DUT is connected, the initial resistance value may appear low due to charging. The resistance value will stabilize over time.
We recommend that you determine the stabilization time specific to your device by observing the resistance value and waiting at least that time before taking measurements.