When unpacked and turned on the POWER switch of TOS series, the PROTECTION message lights on the front panel. What should I do?

The interlock is activated for the safety.
Connect the included SIGNAL I/O plug to the SIGNAL I/O connector to temporarily release the interlock.
Use the interlock function for the actual tests by following the user’s manual.