When using TOS9200 series to measure the insulation resistance of power filter boards, what causes the readings to fluctuate?

This is probably due to the charging of the capacitor in the filter circuit under test.
The higher value obtained after charging is considered the correct measurement result.
Please check if there are any capacitive components such as capacitors in the circuit to be measured.