AC voltage measurement (ACV)
| Item | DME1700 | DME1701 |
|---|
| Measurement range | Resolution | Frequency | Accuracy (1 year) | Temperature coefficient/°C | Accuracy (1 year) | Temperature coefficient/°C |
|---|
| 100 mV to 750 V | 0.1 μV to 1 mV | 3 Hz to 5 Hz | 1.00+0.03 | 0.100+0.003 | 1.00+0.04 | 0.100+0.003 |
|---|
| 5 Hz to 10 Hz | 0.38+0.03 | 0.035+0.003 | 0.35+0.04 | 0.035+0.003 |
|---|
| 10 Hz to 20 kHz | 0.09+0.03 | 0.005+0.003 | 0.06+0.04 | 0.005+0.003 |
|---|
| 20 kHz to 50 kHz | 0.15+0.05 | 0.011+0.005 | 0.12+0.05 | 0.011+0.005 |
|---|
| 50 kHz to 100 kHz | 0.63+0.08 | 0.060+0.008 | 0.60+0.08 | 0.060+0.008 |
|---|
| 100 kHz to 300 kHz | 4.00+0.50 | 0.200+0.020 | 4.00+0.50 | 0.200+0.020 |
|---|