Model | PLZ150U | PLZ70UA | ||
---|---|---|---|---|
Operating range The value inside parentheses is the conductance.*1 | Range | H | OPEN to 50 mΩ (0 S to 20 S) | OPEN to 100 mΩ (0 S to 10 S) |
M | OPEN to 500 mΩ (0 S to 2 S) | OPEN to 1 Ω (0 S to 1 S) | ||
L | OPEN to 5 Ω (0 S to 200 mS) | OPEN to 10 Ω (0 S to 100 mS) | ||
Selectable range | 0 % to 105 % of f.s | |||
Resolution The value inside parentheses is the operating range. | Range | H | 0.2 mS (0 S to 2 S) | 0.1 mS (0 S to 1 S) |
2 mS (2 S to 20 S) | 1 mS (1 S to 10 S) | |||
M | 20 μS (0 S to 200 mS) | 10 μS (0 S to 100 mS) | ||
0.2 mS (200 mS to 2 S) | 0.1 mS (100 mS to 1 S) | |||
L | 2 μS(0 S to 20 mS) | 1 μS(0 S to 10 mS) | ||
20 μS(20 mS to 200 mS) | 10 μS(10 mS to 100 mS) | |||
Accuracy of setting*3 | Range | H, M, and L | ±(0.5 % of set*4 + 0.5 % of f.s*5) + Vin /500 kΩ |
*1 Conductance [S] = (Input current [A]/input voltage [V]) = (1/resistance [Ω]) *2 Conductance f.s *3 Converted value in terms of the input current, during remote sensing *4 set = input voltage × specified conductance = (input voltage/specified resistance) *5 f.s = Rated current of the specified range