Model | PLZ-30F | PLZ-50F | |
---|---|---|---|
Rated supply voltage | 100 VAC to 240 VAC (90 VAC to 250 V) single phase | ||
Rated frequency | 50 Hz or 60 Hz (47 Hz to 63 Hz) | ||
Power consumption | Frame alone | 33 VA or less | 40 VA or less |
When load units are installed in all channels | 300 VAmax | 500 VAmax | |
Cooling method | Forced air cooling using a heat sensing variable speed fan. | ||
Operating temperature range | 0 °C to 40 °C (32 °F to 104 °F) | ||
Operating humidity range | 20 % to 85 % RH (without condensation) | ||
Storage temperature range | –20 °C to 70 °C (-4 °F to 158 °F) | ||
Storage humidity range | 90 % RH or less (without condensation) | ||
Insulation resistance | Between primary and chassis | 500 VDC, 30 MΩ or more (ambient humidity of 70% RH or less) | |
Withstanding voltage | Between primary and chassis | No abnormalities at 1500 VAC for 1 minute. | |
Earth continuity | 25 Aac, 0.1 Ω or less | ||
Battery backup | Backs up the setup data immediately before the power is turned off Battery life: 3 years or longer (at 25 °C) | ||
Number of installable load modules | 3 | 5 | |
Dimensions (mm) | See outline drawing. | ||
Weight | Frame alone | Approx. 5 kg (11.02 lb) | Approx. 7 kg (15.43 lb) |
Accessories | Power cord | 1 pc. (with SVT3, 18AWG, 3-prong plug, cable length of 2.4 m) | |
Blank panel (front panel) | 2 pcs. maximum*1 | 4 pcs. maximum*1 | |
Protection dummy plug | 2 pcs. (for the FRAME CONT connector, attached to the unit) | ||
Operation Manuals | 1 pc. | ||
Electromagnetic compatibility*2*3 | Complies with the requirements of the following directive and standard. EMC Directive 2014/30/EU EN 61326-1 (Class A) EN55011 (Class A*4, Group 1*5) EN 61000-3-2 EN 61000-3-3 | ||
Safety | Complies with the requirements of the following directive and standard. Low Voltage Directive 2014/35/EU*3 EN 61010-1 (Class I*6, Pollution degree 2*7) |
*1 In products that have load units installed, blank panels are installed in the empty slots. In products that contain the frame alone, the maximum number of blank panels are installed. *2 Does not apply to specially ordered or modified PLZ-U *3 Limited to products that have the CE mark/UKCA mark on their panels. *4 This is a Class A equipment. This product is intended for use in an industrial environment. This product may cause interference if used in residential areas. Such use must be avoided unless the user takes special measures to reduce electromagnetic emissions to prevent interference to the reception of radio and television broadcasts. *5 This is a Group 1 equipment. This product does not generate and/or use intentionally radio-frequency energy, in the from of electromagnetic radiation, inductive and/or capacitive coupling, for the treatment of material or inspection/analysis purpose. *6 This is a Class I equipment. Be sure to ground the this product’s protective conductor terminal. The safety of this product is only guaranteed when the product is properly grounded. *7 Pollution is addition of foreign matter (solid, liquid or gaseous) that may produce a reduction of dielectric strength or surface resistivity. Pollution degree 2 assumes that only non-conductive pollution will occur except for an occasional temporary conductivity caused by condensation.