Overvoltage protection (OVP) | Turns off the load at 110 % of the rated voltage |
---|---|
Overcurrent protection (OCP) | Set the value in the range of 0 % to 110 % of the rated current of H range. Trips at the value or 110% of the rated current of the range, whichever is less. The action taken when the OCP trips can be set to load off or limit. |
Overpower protection (OPP) | Set the value in the range of 0 % to 110 % of the rated power of H range. Trips at the value or 110% of the rated power of the range, whichever is less. The action taken when the OPP trips can be set to load off or limit. |
Overheat protection (OHP) | Trips when the heat sink temperature reaches 95 °C. The action taken when the OHP trips is to turn the load off. |
Reverse connection protection (RVP) | Short-term protection provided by a short-circuit system using a protection diode. The action taken when the OHP trips is to turn the load off. |
Undervoltage protection (UVP) | Set the value to off or in the range of 0 % to 100 % of the rated voltage. The action taken when the OHP trips is to turn the load off. |
27805
PLZ-U_Protection function
Related specifications : PLZ-U Series
- PLZ-U_Diverse load on/off operations
- PLZ-U_Load Unit
- PLZ-U_Frame section (for page)
- PLZ-U_Switching function
- PLZ-U_Sequence setting parameters
- PLZ-U_Parallel operation
- PLZ-U_Channel-synchronized operations
- PLZ-U_High precision and high resolution
- PLZ-U_Variable slew rate
- PLZ-U_Frame Details
- PLZ-U_CR mode
- PLZ-U_General specifications
- PLZ-U_Communication function
- PLZ-U_External analog control
- PLZ-U_Inter-frame control and external control
- PLZ-U_Remote sensing & Other
- PLZ-U_Protection function
- PLZ-U_Sequence function
- PLZ-U_Soft start
- PLZ-U_Slew rate
- PLZ-U_Switching mode
- PLZ-U_Wattmeter
- PLZ-U_Ammeter
- PLZ-U_Voltmeter
- PLZ-U_CV mode
- PLZ-U_CC mode
- PLZ-U_Rating
- PLZ-U_Frame
- PLZ-U_Load Unit