PMX-E (2026/05)_spec

ModelPMX18-2EPMX18-5EPMX35-1EPMX35-3EPMX70-1EPMX110-0.6EPMX250-0.25EPMX350-0.2EPMX500-0.1E
AC input
Rated input rating100 VAC, 50 Hz / 60 Hz, single phase (117 / 200 / 217 / 234 VAC available as factory options)
Input voltage range± 10 %
Input frequency range47 Hz to 63 Hz
Inrush current (MAX)50 A60 A45 A60 A65 A55 A40 A55 A40 A
Power150 VA310 VA150 VA310 VA230 VA210 VA210 VA230 VA170 VA
Output
RatingOutput voltage18.00 V18.00 V35.00 V35.00 V70.00 V110.0 V250.0 V350.0 V500.0 V
Output current2.000 A5.000 A1.000 A3.000 A1.000 A0.600 A0.250 A0.200 A0.100 A
Output power36 W90 W35 W105 W70 W66 W62.5 W70 W50 W
VoltageSetting range0 V to 18.90 V0 V to 18.90 V0 V to 36.75 V0 V to 36.75 V0 V to 73.50 V0 V to 115.5 V0 V to 262.5 V0 V to 367.5 V0 V to 525.0 V
Setting resolution *11 mV2 mV10 mV
Setting accuracy± (0.2 % of setting + 0.1 % of rating)
Power fluctuation*2±1 mV±1 mV±3 mV±3 mV±5 mV±7 mV±15 mV±25 mV±30 mV
Load variation*3±2 mV±5 mV±3 mV±4 mV±5 mV±7 mV±15 mV±25 mV±30 mV
Transient response *450 μs100 μs
Ripple noise (rms) *50.5 mV (1 mV) *60.5 mV0.5 mV (1 mV) *60.5 mV1 mV2 mV3 mV5 mV10 mV
Rise time*7Rated load≤ 120 ms (≤ 20 ms) *6≤ 120 ms≤ 120 ms (≤ 20 ms) *6≤ 120 ms≤ 150 ms≤ 120 ms≤ 120 ms≤ 150 ms≤ 120 ms
No load≤ 120 ms (≤ 20 ms) *6≤ 120 ms≤ 120 ms (≤ 20 ms) *6≤ 120 ms≤ 150 ms≤ 120 ms≤ 120 ms≤ 150 ms≤ 120 ms
Fall time *8Rated load≤ 50 ms (≤ 20 ms) *6≤ 50 ms≤ 50 ms (≤ 20 ms) *6≤ 50 ms≤ 50 ms≤ 50 ms≤ 50 ms≤ 80 ms≤ 50 ms
No load≤ 120 ms (≤ 20 ms) *6≤ 120 ms≤ 120 ms (≤ 20 ms) *6≤ 120 ms≤ 150 ms≤ 120 ms≤ 120 ms≤ 150 ms≤ 60 ms
Max. remote sensing compensation voltage0.6 V (one‑way)
Temperature coefficient (TYP)100 ppm/℃
Current
Setting range0 A to 2.10 A0 A to 5.25 A0 A to 1.05 A0 A to 3.15 A0 A to 1.050 A0 A to 0.630 A0 A to 0.262 A0 A to 0.210 A0 A to 0.105 A
Setting resolution *10.1 mA
Setting accuracy± (0.3 % of setting + 0.1 % of rating)
Power fluctuation*2±5 mA±2 mA±2 mA±1 mA±1 mA±1 mA
Load variation*3±5 mA±5 mA±5 mA±5 mA±5 mA±3 mA
Ripple noise (rms) *51 mA (2 mA) *62 mA1 mA (2 mA) *61 mA1 mA
Rise time *7 *9Rated load- (≤ 20 ms) *6- (≤ 20 ms) *6
Fall time *8 *9Rated load- (≤ 20 ms) *6- (≤ 20 ms) *6
Temperature coefficient (TYP)200 ppm/°C

*1.You can set a value down to 1/10 of the minimum digit by using the SHIFT key and knob together or with a communication command. *2.100 Vac to 90 Vac or 100 Vac to 110 Vac, fixed load. *3.For 0 % to 100 % of output voltage or output current. *4.The amount of time required for the output voltage to return to a value within “rated output voltage ± (0.05 % + 10 mV).” When the load current is changed from 10 % to 100 % of the rated output current. *5.At rated load. When the measurement frequency bandwidth is 10 Hz to 1 MHz. *6.The values ​​in parentheses apply to models equipped with the high-speed rising/falling function (factory option). *7.The time for the output voltage to rise from 10 % to 90 % of the rating when the output is turned on. *8.The time for the output voltage to fall from 90 % to 10 % of the rating when the output is turned off. *9.The current rise/fall time specifications apply only to models equipped with the high-speed rising/falling function.