PWR-01_【Common specifications】Signal output and input

Item/Model400W800W1200W2000W
Monitor signal outputVoltage monitor (VMON)Selectable monitor voltage range: 0 V to 5 V or 0 V to 10 V
Accuracy2.5 % of f.s. *1
Current monitor (IMON)Selectable monitor voltage range: 0 V to 5 V or 0 V to 10 V
Accuracy2.5 % of f.s. *1
Status signal output *2OUTON STATUSOn when output is on.
CV STATUSTurns on during CV operation
CC STATUSTurns on during CC operation
ALARM STATUSTurns on when an alarm has been activated
POWER ON STATUSTurns on when the power is turned on
Trigger signalInput (TRG IN)Logic selectable: LOW (0 V to 1.5 V), HIGH (3.5 V to 5 V)
Input impedance: 10 kΩ (TYP)
Output (TRG OUT)Logic selectable: LOW (0 V to 0.6 V), HIGH (4.2 V to 5 V)
Pulse width: 100 μs (TYP)

*1. f.s. is the full scale at the selected range. It is 10 V for the 10 V range and 5 V for the 5 V range. *2. Photocoupler open collector output; maximum voltage 30 V, maximum current (sink) 8 mA; isolated from the output and control circuits; status commons are floating (withstand voltage of less than or equal to 60 V); and status signals are not mutually isolated.