TOS5400_Insulation Resistance Test Section-Judgment function

ItemTOS5402TOS5403
Judgment method and judgment operation*1FunctionThe output is shut off when a judgment is made. Beep volume level can be set for PASS and FAIL separately.
UPPER FAILJudgment methodUPPER FAIL results when a resistance greater than or equal to the Upper limit is detected. Judgment is not made during or Voltage rise time.
DisplayShown on the display.
Beep soundOn
SIGNAL I/OThe U FAIL signal is generated continuously until a STOP signal is received.
LOWER FAILJudgment methodLower limit (Lower) is detected or when there is an abnormality during the voltage rise.
DisplayShown on the display.
Beep soundOn
SIGNAL I/OThe L-FAIL signal is generated continuously until a STOP signal is received.
PASSJudgment methodPASS judgment is made if Upper-FAIL or Lower-FAIL has not occurred when the test time elapses.
DisplayShown on the display.
Beep soundON (fixed at 0.2 s regardless of Pass Hold setting)
SIGNAL I/OThe PASS signal is generated for the length of time specified by the Pass Hold setting. If Pass Hold is set to Infinity, the PASS signal is generated continuously until a STOP signal is received.
Upper limit setting range0.001MΩ to 10.000GΩ
Lower limit setting range0.000MΩ to 9.999GΩ
Judgment accuracy (common between Upper and Lower)*2 *3 *4Measurement accuracy + 2 digits

*1 In the following conditions, charging current may activate a FAIL judgment or the protection function: EUT capacitance: 2μF; Rise Time: 1.5 seconds or less; Judge Delay: 2.0 seconds or less; Low Pass Filter: OFF
*2 Humidity: 70%rh or less (no condensation), when there is no interference caused by wobbly test leads or other problems.
*3 For judgments of 5μA or less, a test time of at least 1.0 seconds is necessary.
*4 When the low pass filter is set to medium, a test time of at least 0.3 seconds is required. When the low pass filter is set to slow, a test time of at least 0.5 seconds is required.