TOS5400_Insulation Resistance Test Section-Output functionÄb0

ItemTOS5402TOS5403
Output voltage25Vdc to 1000Vdc, negative polarity
Output setting accuracy±(1.0% of setting + 2V)
Output setting range0.00kV to 1.02kV
Output setting regulation1V
Max. rated load1W (-1000Vdc/1mA)
Max. rated current1mA
Ripple1000V no load2Vp-p or less
Max. rated load10Vp-p or less
Voltage regulation1 % or less (when changing from maximum rated load to no load)
Short-circuit current12mA or less
Discharge functionForced discharge after test completion (discharge resistance: approx. 25kΩ)*1
Maximum capacitance: 2μF*2
Discharge time: 0.0 to 100.0 seconds
Discharge function during interlock activation: ON/OFF
Output voltage monitor functionIf the output voltage exceeds ±(10% of setting + 10V), the output is turned off, and the protection function is activated.

*1 In the following conditions, charging current may activate a FAIL judgment or the protection function: EUT capacitance: 2μF; Rise Time: seconds or less; Judge Delay: 2.0 seconds or less; Low Pass Filter: OFF
*2 Maximum capacitance that can be discharged within 2 seconds in a insulation resistance test. Tests can still be conducted beyond the maximum capacitance, but the discharge time will increase.